Customization: | Available |
---|---|
After-sales Service: | on Site Installation and Online Support |
Warranty: | 12 Months |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Product Model | KS-VGB-6 vacuum glove box |
Chamber Size | 2440 × 780 × 900 mm |
Material | SUS 304 stainless steel, 3 mm thick |
Glove Ports | Ø220 mm, anti-corrosion aluminum alloy |
Gloves | PIERCAN butyl rubber gloves, 8", 0.4 mm thick |
Front Window | 10 mm tempered glass, inclined design with high-sealing gasket, front-opening structure |
Working Pressure Range | -12 mbar to +12 mbar |
Lighting | Energy-saving LED with anti-reflection coating |
Internal Power Strip | 220V ±10%, 50Hz ±10%, multi-socket |
Rear Ports | 5 × KF40 standard interfaces |
Shelves | Two adjustable stainless steel shelves |
Support Frame | 914 mm height, with leveling, mobility, and lockable casters |
Leak Rate | ≤0.05 vol%/h (ISO 10648-2 compliant) |
Large Antechamber Shape | Cylindrical, SUS 304 stainless steel |
Large Antechamber Size | Ø360 mm × 600 mm, with sliding tray |
Large Antechamber Tray | 230 mm × 595 mm, extendable depth 361 mm |
Large Antechamber Vacuum | ≤ -0.1 MPa, vacuum gauge |
Small Antechamber Shape | Cylindrical, SUS 304 stainless steel |
Small Antechamber Size | Ø155 mm × 320 mm, manual clamping handle |
Small Antechamber Vacuum | ≤ -0.1 MPa, vacuum gauge |
Control System | Siemens LCD touchscreen, Chinese/English menu |
Control Features | Auto diagnosis, auto restart, pressure and vacuum control, password protection |
Foot Pedal | Pressure control for easy operation |
Purification Capability | H2O and O2 <1 ppm |
Control Method | PLC touchscreen for heating, vacuuming, purification, regeneration |
Working Gas | N2or Ar (recycling mode) |
Circulation Fan | Elektror VFD fan, 0-80 m³/h |
Purification Columns | (5 kg O2 absorbent, 60L), UOP (5 kg desiccant, 2 kg active) |
Purification Valves | Fully automatic pneumatic valves, modular design |
Filter | HEPA filter, 0.3 μm pore size |
Regeneration Gas | 5-10% H2 mixed with N2 or Ar |
Vacuum System | Edwards RV12 vacuum pump (UK), optional, independent control for chamber and antechambers |